The CR-100 is a rugged, sensitive, fast colorimeter that is powered directly from the connected USB port.
Calman Required Version
- Version 5.4.0 or later
USB Driver Installation
The CRI meter USB driver should be installed before the meter is connected to a computer USB port for the first time. Available at the Colorimetry Research website
Windows 7
- Install the CRI meter USB driver, available for download from the Colorimetry Research website
Windows 8 or Windows 10
The Colorimetry Research USB driver is unsigned, requiring the following procedure for Windows 8 or Windows 10
Windows 8
- Open Settings (move your mouse to the top or bottom right corner of the screen and wait for the charm bar to appear, then click the Gear icon)
- Click Change PC Settings.
- Click General.
- Scroll down and click Restart Now under Advanced Startup.
- Click Troubleshoot.
- Click Advanced Options.
- Click Windows Startup Settings.
- Click Restart.
- When the computer restarts, select Disable Driver Signature Enforcement from the list.
You can now load the unsigned driver(s).
After restarting the computer, any drivers that were installed will continue to function properly.
Note: This may need to be repeated each time you update your graphics driver
Windows 8.1 and Windows 10
- While holding down the Shift key, go to SettingsPowerRestart
- On shutdown, you will be presented a list of Options. Select Troubleshoot.
- On the Troubleshoot menu, select Advanced Options.
- On Advanced Options, select Startup Settings.
- Click Restart – Machine will reboot.
- This will bring you to Startup Settings. Select 7 (Disable driver signature enforcement).
- When Windows starts, Driver Signature Enforcement is disabled.
You can now load the unsigned driver(s).
After a normal reboot, Driver Signature Enforcement will re-enable. All drivers installed while it was disabled will continue to function properly.
To check the USB driver installation:
- Connect the CRI meter to a USB port on the computer.
Open Device Manager from the Control Panel (or click the Start button and type “device manager”).
On the Device Manager window, check that, under Ports (COM & LPT), there is a listing for Colorimetry Research, Inc. Meter (COMx).
Calibration Matrices
Calman is able to access any meter calibration matrices in the CRI meter’s internal user memory. As supplied from the factory, there are calibration matrices in the factory memory, but none in the user memory. Before using the meter with Calman, you need to copy any of the calibration matrices that you wish to use from the factory memory to the user memory. You need to check this by running the Colorimeter Utility, supplied by CRI.
To copy calibration matrices from the meter’s factory memory to the user memory:
- In the Colorimeter Utility software, on the Calibration tab, click the Select Calibration Matrix button, to the lower right
- On the Select Calibration Matrix dialog that opens, select any or all of the calibration matrices that you may wish to use with the CRI meter.
- Click OK.
- The Calibration tab should now show the calibration matrices that you selected
Calman Meter Connection Procedure
- On the Meter Settings tab, click Find Meter.
- On the Find Meters popup dialog:
- Leave Com Port set to auto, or, select the com port that was assigned to the meter (shown in Device Manager, under Ports (com &LPT)).
- Check the box for Colorimetry Research Inc Meters (USB)
- Click Search
- On the Meter Settings tab:
- Select the Meter Mode for your target display type.
- Select the Sync Mode that produces the most consistent measurements on the target display (check with Read Continuous on 100% White).
- Off: The default sync mode for the CRI meter. This is the fastest mode of meter operation, but may not provide consistent measurements on some display types.
- On: Takes one sync measurement to synchronize the meter with the display refresh rate, then performs all of the series or continuous data measurements. This is only slightly slower, but may not provide optimum sync with displays that have different refresh rates at lower luminance levels.
- Auto: Takes a sync measurement prior to each data measurement. This is slower yet, but provides optimum sync with all display types, at all luminance levels.
- If you wish to use the extended exposure settings available in the CRI measurement application, make your settings in the CRI app. Then, on the Calman Meter Settings tab, select User Mode for the Calman exposure mode. Your main use of this capability might be to take have the CR-100 take multiple readings and average them (setting the Exposure Multiplier) before sending the results to Calman, as this is much faster than having Calman take multiple averaged readings. This might be done to improve the continuous reading stability on displays such as plasma.